Respironics Trilogy 202

Ventilator

Product number:

NOCTN98

The Trilogy 202 is both a volume-control & pressure-control ventilator for invasive and noninvasive ventilation. Versatile breath delivery and setup options provide greater continuity of care.

Specifications Show all

Patient types
Pediatric
≥ 5 kg or greater
Adult
Yes
Volume Modes
MPV Mouthpiece Ventilation Mode
Yes
Volume modes
Pressure control (PC)
Yes
Synchronized intermittent mandatory ventilation (SIMV)
Yes
SIMV with pressure support (SIMV w/PS)
Yes
Control ventilation (CV)
Yes
Pressure modes
Pressure control (PC)
Yes
Pressure control-SIMV (PC-SIMV)
Yes
Spontaneous ventilation (S)
Yes
Spontaneous ventilation with timed back-up (S/T)
Yes
Timed ventilation (T)
Yes
Continuous positive airway pressure (CPAP)
Yes
Average volume assured pressure support (AVAPS)
With passive circuit type, S, S/T, PC, and T modes only
Synchrony features
Auto-Trak sensitivity
Auto-Trak sensitivity Auto-adaptive triggering, cycling, and leak compensation (available in all modes, passive circuit only)
Adjustable flow triggering
1 – 9 l/min (available in all modes and circuit types) l/min
Circuit types
Active exhalation valve with proximal airway pressure (PAP)
Yes
Active exhalation valve with flow sensor
Yes
Passive exhalation port
Yes
Controls
IPAP
50 cmH20
EPAP/PEEP (a)
0 - 25 cmH20 (active valve circuits) cmH20
EPAP/PEEP (b)
4 - 25 cmH20 (passive leak port circuits) cmH20
CPAP
4 - 20 cmH20 (passive leak port circuits) cmH20
Pressure support
0 – 30 cmH2O cmH20
Tidal volume
50 - 2000 ml ml
Breath rate(a)
0 - 60 BPM (AC mode) BPM
Breath rate(b)
1 - 60 BPM (all other modes) BPM
Inspiratory time
0.3 - 5.0 s Synchrony features
Rise time
1 - 6 (relative scale)
Ramp start pressure (a)
0 - 25 cmH20 (active circuits) cmH20
Ramp start pressure (b)
4 - 25 cmH20 (passive circuits) cmH20
Ramp start pressure (c)
4 - 19 cmH20 (CPAP mode) cmH20
Ramp time Off
5 - 45 min min
C-Flex Off
1 - 3 (relative scale)
Flow trigger sensitivity
1 - 9 l/min l/min
Flow cycle sensitivity
10 - 90% %
Monitored parameters
Tidal volume
0 - 2000 ml ml
Minute ventilation
0 - 99 l/min
Estimated leak rate
0 - 200 l/min
Respiratory rate
0 - 80 BPM 0 - 80 BPM BPM
Peak inspiratory flow
0 - 200 l/min
Peak inspiratory pressure
0 - 99 cmH20
Mean airway pressure
0 - 99 cmH20
% patient triggered breaths
0 - 100 %
I:E ratio
9.9:1 - 1:9.9
Alarms
Circuit disconnect Off
10 - 60 s s
Apnea Off
10 - 60 s and 4 - 60 BPM s/BPM
High tidal volume Off
50 - 2000 ml ml
Low tidal volume Off
50 - 2000 ml ml
High minute ventilation Off
1 - 99 l/min Off, 1 - 99 l/min l/min
Low minute ventilation Off
1 - 99 l/min l/min
High respiratory rate Off
4 - 80 BPM BPM
Low respiratory rate Off
4 - 80 BPM BPM
Oxygen
FiO2
21 - 100% %
02 flush
2 min at 100%
02 input pressure range
276 - 600 kPa (40 - 87 psi) kPa/psi
Environmental
Operating temperature
5 – 40°C °C
Storage temperature
-20 – 60°C °C
Relative humidity
15 - 95% %
Atmospheric pressure
60 - 110 kPa (450 - 825 mmHg) kPa/mmHg
Electrical
Input voltage
100 - 240 VAC and 50/60 Hz and 2.1 A VAC/Hz
Detachable battery voltage
14.4 VDC
Internal battery life
3 hours under normal conditions hours
Detachable battery life
3 hours under normal conditions hours
external battery connection
12 VDC
Compliance
IEC 60601-1 Medical electrical equipment Part 1: General requirements for safety
Yes
IEC 60601-1-2 General requirements for safety – collateral standard Electromagnetic compatibility – requirements and tests
Yes
IEC 60601-2-12 Medical electrical equipment Part 2-12: Particular requirements for the safety of lung ventilators – Critical Care ventilators
Yes
Options
Detachable back-up battery
Offering up to 3 hours additional operating time
Hospital roll stand
Provides convenient accessory basket and humidifier mount
DirectView
1 GB SD card data storage integrated into the ventilator software