Respironics V60

Non-invasive ventilator

Product number:

NOCTN96

Because every breath matters and each one is different, patients need a ventilator that follows their breathing pattern, whatever their acuity. The Philips Respironics V60 uses auto-adaptive technology to provide patient synchrony and therapy acceptance. Improved signal processing technology is finely tuned for both adult and pediatric patients.

Specifications Show all

Patient types
Suitability
Pediatric (≥ 20 kg) and Adult
Settings
C-Flex
Off, 1 to 3
CPAP
4 to 25 cm H2O
EPAP
4 to 25 cm H2O
IPAP
4 to 40 cm H2O
Inspiratory time
0.30 to 3.00 sec
Max P (AVAPS maximum IPAP)
6 to 40 cm H2O
Min P (AVAPS minimum IPAP)
5 to 30 cm H2O
Oxygen percentage
21 to 100 %
Ramp time
Off, 5 to 45 min
Respiratory rate
4 to 60 BPM
Rise time
1 to 5
Triggering and cycling
Auto-adaptive (Auto-Trak) or Auto-Trak Plus (optional). Trigger: normal, +1 to +7. Cycle: -2, -1, normal, +1 to +6
AVAPS target tidal volume
200 to 2000 ml BTPS
Max E
0 to 100 cm H2O/l
Max R
0 to 50 cm H2O/l/s
PPV%
0 to 100 %
Max P (PPV maximum pressure limit)
5 to 40 cm H2O
Max V (PPV maximum volume limit)
200 to 3500 ml
Patient Data
Breath phase/trigger indicator
Spontaneous, Timed, Exhale
PIP
0 to 50 cm H2O
Patient/total leak
0 to 200 l/min BTPS
Patient Trigger
0 to 100 %
Respiratory Rate
0 to 90 BPM
Ti/Ttot
0 to 91 %
Minute volume
0 to 99.0 l/min BTPS
Tidal volume
0 to 3000 ml BTPS
General
Oxygen inlet pressure range
276 to 600 kPa (40 to 87 psig) kPa (psig)
Weight with optional battery
10.9 kg (24 lb) kg (lb)
Weight without optional battery
10.0 kg (22 lb) kg (lb)
Dimensions
33.7 cm H x 39.4 cm W x 42.9 cm D (13.3"H x 15.5" W x 16.5" D) cm (inch)
Electrical
AC voltage
100 to 240 VAC
AC frequency
50 to 60 Hz
AC power
300 VA
Battery operating time
6 hours in normal conditions
Waveform window
Pressure waveform
0 - 50 cmHzO
Flow waveform
240-2401 /min btps
Volume waveform
0 - 3,500 ml btps
Alarm adjustable range
Hi Rate (hi respiratory rate alarm)
5 - 90 bpm
Lo Rate (low respiratory rate alarm)
1 - 89 bpm
Hi Vr (high tidal volume alarm)
200 - 3,500 ml
Lo Vr (low tidal volume alarm)
OFF - 1,500 ml
HIP (high inspiratory pressure alarm)
5 - 50 cmHzO
LIP (low inspiratory pressure alarm)
OFF, 1 - 40 cmHzO
Lo VE (low minute ventilation alarm)
OFF, 0.1 - 991 /min
LIP T (low inspiratory pressure delay time)
5 - 60 sec